YOKOGAWA vigilantplant OX102 User Manual page 15

Current limit type oxygen analyzer
Table of Contents

Advertisement

IM 11M10A01-03E
(15) Fine tune mode
(16) Operating temperature : Converter: 0 to 50 C
(17) Extension cable : 4 m, 9 m, 29 m, operating temperature: 0 to 70 C
(18) Gas flow rate
: 1 m/sec max.
(19) Operating pressure
(20) Power supply
: 100 to 240 V AC, 50/60 Hz, approx. 15 W
(21) Storage temperature
(22) Self-diagnostic function : Detects sensor disconnection
(23) Sensor replacement
: The measuring value can be adjusted to true analyzer
value.
Sensor
0 to 150 C
0 to 250 C within 50mm of sensor tip
: 1013
40 hPa abs
: 0 to 70 C for both sensor and converter
: Compatibility can be achieved by entering a sensor-
specific constant
1. Overview
1-3

Advertisement

Table of Contents