Troubleshooting
52
Table 11: Error Codes (Continued)
Error
Meaning
Code
518
Application program is corrupt
519
Invalid FE102 version
520
Error in the start up sequence
521
OS multitasking service failure
522
A state machine has received an unknown state transition
523
The operation just attempted was not completed successfully
example, Institutional Defaults could not be reset
524
An unexpected value was received
parameter was passed to a function
525
EEPROM CRC failure
526
SpO
module not responded
2
527
Institutional parameters lost
section CRC corrupt
528
Current settings lost
corrupt
531
External watchdog failure
532
Power PC watchdog failure
533
Boot NVROM uninitialized error
534
Failed CRC check of application code in flash
535
Failed periodic ram CRC check on application code running in RAM
562
SpO
front end reset
2
563
SpO
reported error
2
564
Clinical mode was exited after input was received
565
Communication failures between software modules
566
Excessive resets before UIF runs
567
An unexpected interrupt has been asserted
568
General failure in UIF module generic post
569
BOOT application program is corrupt
570
RTC was restarted
574
Excessive restarts within 1 minute
-
for example, an out
-
e.g. for UIF: Institutional EEPROM
-
e.g. for UIF: Institutional EEPROM section CRC
-
CRC does not match
-
for
-
-
of
range