Performance Verification
24
3. The N-550B will display:
•
75 %SpO2 (test pass criteria is 73 to 77 %SpO
•
60 bpm (test pass criteria is 57 to 63 bpm)
•
alarm
Pulse Amplitude indicator - high level modulation
•
4. Perform Test #1: BPM on page 21. The Pulse Amplitude indicator
should indicate high level modulation.
5. Perform Test #2: SpO2 on page 22. The Pulse Amplitude indicator
should indicate high level modulation.
6. Press the SRC-MAX % Modulation selection button. The SRC-MAX %
Modulation
LED will light.
7. The N-550B Pulse Amplitude indicator will stabilize and illuminate 4
bars with each pulse beat.
8. The N-550B will display:
•
75 %SpO2
•
60 bpm
•
alarm
Pulse Amplitude indicator - low level modulation
•
)
2