Disassembly Guide
Main PCB Removal
60

Figure 22: MP-506 Cable Connections

4. Disconnect the main PCB cable connector from the MP-506 PCB
connector, item 2.
5. Carefully lift the MP-506 PCB out of the top case. See Figure 23.

Figure 23: MP-506 Removal

1. Perform the procedure Prior to Disassembly on page 54.
2. Perform the procedure Separating Top and Bottom Cases on page 54.