Figure 44: Data Port Pin Layout - Nellcor OxiMax N-550B Service Manual

Table of Contents

Advertisement

Data Port Interface Protocol
98
Table 19: Data Port Pinouts (Continued)
Pin
10
Signal Ground
11
Nurse Call (RS-232 level output)
12
TXD- (RS-422 negative output) (-5 to -12 VDC with no audible
alarm, 5 to 12 VDC with audible alarm)
13
NC
14
NC
15
Nurse Call Common for Dry Contacts
The pin layouts are illustrated in Figure 44. The conductive shell is used as
earth ground. An AMP connector is used to connect to the data port. Use
AMP connector (AMP P/N 747538-1), ferrule (AMP P/N 1-747579-2) and
compatible pins (AMP P/N 66570-2). The serial cable must have a braided
shield providing 100% coverage, such as Beldon cable (Beldon P/N 9616) or
equivalent. Connectors at both ends of the serial cable must have the shield
terminated to the full 360 degrees of the connector's metal shell.
CAUTION: Do not create sharp bends in the cable; this may tear or
break the shield.

Figure 44: Data Port Pin Layout

Pins 2, 3, and 5 provide data in RS-232 format.
Pins 1, 4, 9, and 12 provide data in RS-422 format.