Service Menu Options
Menu Item 6 (In-Sensor Trend Mode)
Menu Item 7 (RS-232 Nurse Call Polarity)
N-550B
When the desired option is indicated in the %SpO
SatSeconds Alarm Limit button to initiate the current selection.
Note: Menu items greater than 5 cannot be accessed when a valid sensor is
connected to the N-550B.
Service menu options can be accessed only when the sensor is disconnected
from the N-550B. Only qualified service personnel should access these
options. Refer to Accessing Menu Items on page 33 for instructions on how to
access the menu options and make selections within them.
This menu item is used to set up the type of event data that a single use
O
M
sensor will record within the sensor. Three options can be accessed
XI
AX
by first pressing the SatSeconds Alarm Limit button when menu item 6 is
displayed. This function is only applicable to O
any event data recorded in the sensor. If the O
recorded in it, the N-550B will record event data in the same format as the
recorded data.
Use the Adjust Up or Adjust Down button to select the desired option.
Option 0 = record SpO
Option 1 = record SpO
Option 2 = do not use in-sensor record function
When the desired option is indicated in the %SpO
SatSeconds Alarm Limit button to initiate the current selection.
This menu item is used to setup the polarity of the RS-232 Nurse Call signal
available at the N-550B rear panel data port. Two options can be accessed by
first pressing the SatSeconds Alarm Limit button when menu item 7 is
displayed.
Use the Adjust Up or Adjust Down button to select the desired option.
Audible Alarm Settings and Service Functions
display, press the
2
M
sensors that do not have
XI
AX
M
sensor has event date
XI
AX
event data
2
plus pulse rate event data
2
display, press the
2
35