CombiFlash NextGen Systems
Section 2 Preparation
Figure 2-4 Location of ELSD gas port (Nitrogen input line circled)
2.6 Connect and Route
Drain Lines
2-6
Risk of fire or equipment damage. Failure to connect drain
lines may allow organic solvents to pool in unsafe areas,
creating a potential for dangerous levels of flammable
vapors. Improper draining may damage the instrument's
internal components.
The CombiFlash NextGen system has drain tubes extending
from its back panel. The tubes drain away any liquid spilled on
the top shelf and the tray beneath the fraction collection racks.
1. Test the fraction collector drain by connecting a vacuum or
air supply source to the outlet end of the drain tube. Then,
verify the presence of such vacuum or air supply source on
the drain (Figure 2-5).
2. Test the top shelf drain by connecting a vacuum or air sup-
ply source to the outlet end of the drain tube. Then, verify
the presence of such vacuum or air supply source on the
drain (Figure 2-6).
EAR99 Technology Technology Subject to Restrictions Contained on the Cover Page
WARNING