3.3.11.25 [Copies]
It is used to set the number of printed paper.
3.3.11.26 [Print curve]
Curve of current printing interface
3.3.11.27 [Print screen]
It is used for printing in full screen.
3.3.11.28 [Display ]
Display of control screen
3.3.11.29 [Full screen]
Screen is switched to full screen. Menus on the right of
screen and parameter state on the left of screen are blanked
for viewing elaborate trace information. Re-press this key
to exit full-screen mode.
3.3.11.30 [Window zoom On Off]
If window zoon function is on, measured signal can be
displayed as
Scan portion
in the lower part of screen.
60
3.3.11.31 [Display line On Off]
Enable or disable display line, or change its display position.
Display line can be as reference for reading.
3.3.11.32 [Amplitude ruler On off]
Enable or disable amplitude ruler function. When it is on,
the left-most part of screen is displayed, easy for observing
signal power.
3.3.11.33 [Backlight brightness]
Set backlight brightness of spectrum analyzer's LCD screen.
You can use number keys, knob and direction keys to revise
parameters.
3.3.11.34 [User calibration ]
Soft menu about user calibration pops up.
Notes:
Set signal generator (frequency: 440MHz, power: -20dB),
connect with spectrum analyzer's RF input, press user
calibration key to execute user calibration.
If there is no need to calibrate compensation data by this
user, press [Reset to factory default] key to clear data and
reset to factory state.