3.3.6.14 [Auto]
Set wave detector in standard wave detection mode (default
mode). In this mode, when noise is detected, measurement
results of positive peak value and negative peak value are
alternately displayed, so as to reach display effect similar
to that generated by analogy instrument. Otherwise, only
positive peak value is displayed.
3.3.6.15 [Normal]
In this mode, when noise is detected, positive peak value
and negative peak value are alternately displayed. Otherwise,
only positive peak value is displayed.
3.3.6.16 [Positive peak]
Select positive peak wave detection mode. In this mode,
wave detector selects the maximum value in sampling data
segment to be displayed in corresponding pixel. Positive
peak wave detector is selected when it is in [Max. hold]
mode .
3.3.6.17 [Negative peak]
Select negative peak wave detection mode. In this mode,
wave detector selects the minimumvalue in sampling data
segment to be displayed in corresponding pixel.
3.3.6.18 [Sampling]
Set wave detector in sampling wave detection mode. This
mode is usually used in video average and noise frequency
marker functions.
3.3.7 MEAS
Provide various advanced measurement functions, including
time spectrum, adjacent channel power, channel bandwidth,
and occupied bandwidth.
3.3.7.1 [Measurement off]
Directly disable currently running measurement functions,
or select to get it disabled in this measurement menu.
3.3.7.2 [Time spectrum On Off]
Enable or disable measurement function of time spectrum.
3.3.7.3 [Adjacent channel power On Off]
Enable or disable adjacent channel power measurement.
After pressing [Measurement setup], soft menu about
parameter setting for adjacent channel power measurement
45