Site Requirements - Agilent Technologies G1314B User Manual

Agilent 1260 infinity variable wavelength detector
Table of Contents

Advertisement

2
Site Requirements and Specifications

Site Requirements

Site Requirements
Site Requirements
A suitable environment is important to ensure optimal performance of the
instrument.
Power Consideration
The detector power supply has wide ranging capabilities, see
Specifications"
mentioned range. Consequently, there is no voltage selector in the rear of
the detector. There are also no externally accessible fuses, because
automatic electronic fuses are implemented in the power supply.
Instrument is partially energized when switched off
WA R N I N G
The power supply still uses some power, even when the power switch on the front
panel is turned OFF. Repair work at the detector can lead to personal injuries, e. g.
shock hazard, when the detector cover is opened and the instrument is connected to
power.
Hazard of electrical shock or damage of your instrumentation
WA R N I N G
can result, if the devices are connected to a line voltage higher than specified.
20
on page 24. It accepts any line voltage in the above
To disconnect the detector from the power line, unplug the power cord.
Connect your instrument to the specified line voltage only.
Agilent 1260 Infinity VWD User Manual
"Physical

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

G1314c

Table of Contents