8
Test Functions

Dark Current Test

Dark Current Test

The dark- current test measures the leakage current from the sample and
reference circuits. The test is used to check for defective sample or
reference diodes or ADC circuits which may cause non- linearity or
excessive baseline noise. During the test, the lamp is switched off. Next,
the leakage current from both diodes is measured.
Figure 39
112
Dark Current Test (Agilent Lab Advisor)
Agilent 1260 Infinity VWD User Manual