Installation Information on Leak and Waste Handling

The Agilent 1200 Infinity Series has been designed for safe leak and waste
handling. It is important that all security concepts are understood and
instructions are carefully followed.
Toxic, flammable and hazardous solvents, samples and reagents
WA R N I N G
The handling of solvents, samples and reagents can hold health and safety risks.
Recommendations for Solvent Cabinet
N O T E
For details, see the usage guideline for the Agilent 1200 Infinity Series Solvent Cabinets.
Agilent 1260 Infinity VWD User Manual
When working with these substances observe appropriate safety procedures (for
example by wearing goggles, safety gloves and protective clothing) as described in
the material handling and safety data sheet supplied by the vendor, and follow good
laboratory practice.
The volume of substances should be reduced to the minimum required for the
analysis.
Never exceed the maximal permissible volume of solvents (6 L) in the solvent
cabinet.
Do not use bottles that exceed the maximum permissible volume as specified in the
usage guideline for the Agilent 1200 Infinity Series Solvent Cabinets.
Arrange the bottles as specified in the usage guideline for the solvent cabinet.
A printed copy of the guideline has been shipped with the solvent cabinet,
electronic copies are available on the Internet.

Installation Information on Leak and Waste Handling