Overview of Maintenance
The following pages describe maintenance (simple repairs) of the detector
that can be carried out without opening the main cover.
Table 13
Simple Repairs
Procedures
Deuterium lamp
exchange
Flow cell exchange
Cleaning flow cell
parts cleaning or
exchange
Leak sensor drying
Leak handling system
replacement
Agilent 1260 Infinity VWD User Manual
Typical Frequency
If noise and/or drift exceeds your application limits or
lamp does not ignite.
If application requires a different flow cell type.
If leaking or if intensity drops due to contaminated flow
cell windows.
If leak has occurred.
If broken or corroded.
Maintenance
Overview of Maintenance
Notes
A VWD test should be performed after
replacement.
A VWD test should be performed after
replacement.
A pressure tightness test should be
done after repair.
Check for leaks.
Check for leaks.
9