Holmium Oxide Test Failed - Agilent Technologies G1314B User Manual

Agilent 1260 infinity variable wavelength detector
Table of Contents

Advertisement

8
Test Functions
Holmium Oxide Test

Holmium Oxide Test Failed

Probable cause
1
2
3
4
116
Detector not calibrated.
Dirty or defective flow cell.
Dirty or defective holmium oxide filter.
Optical misalignment.
Suggested actions
Recalibrate the detector.
Repeat the test with the flow cell removed. If
the test is OK, exchange the flow cell
components.
Run the holmium oxide filter test. If the test
fails, contact your Agilent service
representative.
Please contact your Agilent service
representative.