Special Settings

The special settings are required for specific actions (normally in a service
case).
Boot-Resident
Firmware update procedures may require this mode in case of firmware
loading errors (main firmware part).
If you use the following switch settings and power the instrument up
again, the instrument firmware stays in the resident mode. It is not
operable as a module. It only uses basic functions of the operating system
for example, for communication. In this mode the main firmware can be
loaded (using update utilities).
Table 25
Boot Resident Settings (without on-board LAN)
Mode Select
TEST/BOOT
Agilent 1260 Infinity VWD User Manual
Setting the 8-bit Configuration Switch (without On-board) LAN
SW1
SW2
SW3
1
1
SW4
SW5
0
0
1
Hardware Information
SW6
SW7
SW8
0
0
12
0
185