6-127. Leveled Sine Wave Flatness Verification

Leveled Sine Wave flatness verification is divided into two frequency bands: 50 kHz to
10 MHz (low frequency) and > 10 MHz to 300 MHz (high frequency). The equipment
setups are different for each band. Leveled Sine Wave flatness is measured relative to 50
kHz. This is determined directly in the low frequency band. The high frequency band
requires a "transfer" measurement be made at 10 MHz to calculate a flatness relative to
50 kHz.

6-128. Equipment Setup for Low Frequency Flatness

All low frequency flatness procedures use the following equipment:
Connect the Calibrator Mainframe SCOPE connector to the 5790A WIDEBAND input
with the BNC(f) to Type N(m) adapter as shown in Figure 6-25. Set the 5790A to
AUTORANGE, digital filter mode to FAST, restart fine, and Hi Res on.
5790A
INPUT 1
1000V RMS MAX
SHELL FLOATING
SHUNT
3V RMS MAX
WIDEBAND
7V RMS MAX
SHELL FLOATING
GROUND
10V PK
MAX
Figure 6-25. Connecting the Calibrator Mainframe to the 5790A AC Measurement Standard

6-129. Equipment Setup for High Frequency Flatness

All high frequency flatness procedures use the following equipment:
When high frequencies at voltages below 63 mV p-p are verified, use the
8481D Power Sensor. Otherwise, use the 8482A Power Sensor.
6-100
5790A/03 AC Measurement Standard with Wideband option
BNC(f) to Type N(m) adapter
BNC cable supplied with the SC300
AC MEASUREMENT
STANDARD
INPUT 2
1000V RMS MAX
SHUNT
INPUT1
INPUT1
INPUT1
INPUT1
HI
2.2 mV
22 mV
220 mV
2.2 V
22 V
.
220 mV
VIEW
REF
6
7
8
9
+/-
DELETE
7 mV
70 mV
700 mV
7V
70V
700V
LO
0
1
2
3
4
5
CLEAR
10V PEAK
MAX
2.2 mV
1kV
AUTO
MAN
ENTER
GUARD
Hewlett-Packard E4418A Power Meter
Hewlett-Packard 8482A and 8481D Power Sensors
BNC(f) to Type N(f) adapter
BNC cable supplied with the Calibrator Mainframe
5500A
CALIBRATOR
NORMAL
AUX
SCOPE
V,
,
A,
-SENSE,
200V PK
RTD
AUX V
MAX
HI
UTIL
SPEC
MENUS
1000V
20V
RMS
RMS
MAX
MAX
TRIG
LO
OUT
1V PK
MAX
20V PK
MAX
POWER
TC
I
20V PK
O
MAX
Note
PREV
STBY
OPR
EARTH
SCOPE
BOOST
MENU
µ
dBm
sec
EDIT
7
8
9
m
V
Hz
SETUP
RESET
FIELD
n
W
¡F
NEW
4
5
6
k
A
¡C
CE
REF
p
MEAS
TRIG
1
2
3
M
F
TC
OUT
MULT
DIV
/ +
0
SHIFT
ENTER
÷
x
POWER
I
O
om034f.eps

Advertisement

Table of Contents
loading

Table of Contents