6-133. Wave Generator Verification

This procedure uses the following equipment:
For wave generation verification procedures, refer to Figure 6-28 for the proper setup
connections.
6-108
Table 6-69. Time Marker Verification
PM 6680Settings
Channel
Filter
A
On
A
On
A
Off
A
Off
A
Off
A
Off
A
Off
A
Off
A
Off
A
Off
A
Off
A
Off
A
Off
C
Off
5790A AC Measurement Standard
BNC(f) to Double Banana adapter
50 Ω feedthrough termination
BNC cable supplied with the Calibrator Mainframe
PM 6680 Reading
PM 6680 Reading
(Frequency)
1
Tolerance
(Period)
24.91E-3 s
4.06E-3 s
3.75E-6 s
900E-09 s
350E-09 s
1.29E-9 s
506E-12 s
251.5E-12 s
25.0E-12 s
1.25E-12 s
500E-15 s
250E-15 s
125E-15 s
50E-15 s