Sc600 Hardware Adjustments; Equipment Required; Adjusting The Leveled Sine Wave Function - Fluke 5500A Service Manual

Multi-product calibrator
Hide thumbs Also See for 5500A:
Table of Contents

Advertisement

5500A
Service Manual
Set the Calibrator Mainframe to SCOPE mode, with the Overload menu on the display.
Connect the BNC cable to the Calibrator Mainframe SCOPE connector. Then follow
these steps to verify the overload function.
1. Connect the 50 Ω feedthrough termination to the end of the BNC cable.
2. Program the Calibrator Mainframe output for 5.000 V, DC (OUT VAL blue
3. Press O on the Calibrator Mainframe to activate the output and verify that the
4. Remove the 50 Ω feedthrough termination before 60 seconds and verify that
5. Reconnect the 50 Ω feedthrough termination to the end of the BNC cable.
6. Program the Calibrator Mainframe output for 5.000 V, AC (OUT VAL blue
7. Press O on the Calibrator Mainframe to activate the output and verify that the
8. Remove the 50 Ω feedthrough termination before 60 seconds and verify that

6-74. SC600 Hardware Adjustments

Hardware adjustments must be made to the leveled sine and edge functions each time the
SC600 is repaired. In addition to the adjustment procedures, this section provides lists of
the required equipment and some recommendations on models that have the capabilities
required by these procedures. Equivalent models can be substituted if necessary.

6-75. Equipment Required

The following equipment is necessary for performing the hardware adjustments
described in this section. The models listed are recommended for providing accurate
results.

6-76. Adjusting the Leveled Sine Wave Function

There are two adjustment procedures that need to be made for the leveled sine wave
function. The first procedure adjusts the balance out of the LO VCO so that the signal is
balanced between the two VCOs. The second procedure adjusts the harmonics.
6-60
softkey), and time limit = 60 s (T LIMIT blue softkey).
OPR display timer increments.
Calibrator Mainframe goes to STBY.
softkey).
OPR display timer increments.
Calibrator Mainframe goes to STBY.
Standard adjustment tool for adjusting the pots and trimmer caps
Extender Card
Oscilloscope Mainframe and Sampling Head (Tektronix 11801 with SD-22/26 or
Tektronix TDS 820 with 8 GHz bandwidth)
10 dB Attenuator (Weinschel 9-10 (SMA), or Weinschel 18W-10, or equivalent)
Cable provided with SC600
Spectrum Analyzer (Hewlett-Packard 8590A)