6-99. Equipment Required for Calibration and Verification

Table 6-47 lists the equipment, recommended models, and minimum specifications
required for each calibration and verification procedure.
Table 6-47. SC300 Calibration and Verification Equipment
Instrument
Wave Generator, Edge Amplitude Calibration, AC Voltage Verification
Digital
HP 3458A
Multimeter
Adapter
Pomona #1269
Termination
BNC Cable
(supplied with SC300)
High-
Tektronix 11801 with
Frequency
Tektronix SD-22/26
Digital Storage
sampling head, or
Oscilloscope
Tektronix TDS 820 with
8 GHz bandwidth
Attenuator
Weinschel 9-10 (SMA)
or Weinschel 18W-10
or equivalent
Adapter
BNC Cable
(supplied with SC300)
Leveled Sine Wave Amplitude Calibration and Verification
AC
Fluke 5790A
Measurement
Standard
Adapter
Pomona #1269
Termination
BNC Cable
(supplied with SC300)
DC and AC Voltage Calibration and Verification, DC Voltage Verification
Digital
HP 3458A
Multimeter
Adapter
Pomona #1269
Termination
BNC Cable
(supplied with SC300)
Model
Voltage
Edge
BNC(f) to Double Banana Plug
Feedthrough 50 Ω ± 1% (used with Edge Amplitude
Calibration and AC Voltage Verification)
Edge Rise Time and Aberrations Verification
Frequency
Resolution
10 dB, 3.5 mm (m/f)
BNC(f) to 3.5 mm(m)
Range
Frequency
BNC(f) to Double Banana Plug
Feedthrough 50 Ω ± 1%
BNC(f) to Double Banana Plug
Feedthrough 50 Ω ± 1%

Equipment Required for Calibration and Verification

Minimum Use Specifications
1.8 mV to ± 105 V p-p Uncertainty: 0.06%
4.5 mV to 2.75 V p-p Uncertainty: 0.06%
2 GHz
4.5 mV to 2.75 V
5 mV p-p to 5.5 V p-p
50 kHz
SC300 Option
6
6-75