3-12. AUX AC Volts

Measure the AUX output using a precision AC Voltmeter. Enter into the 5500A the
measured values of each step listed in Table 3-7 when prompted to do so.
Step
NORMAL Output
1
2
3
4
5
6
7

3-13. Resistance

Use a precision DMM to measure the resistance output. Figure 3-3 shows the four-wire
connections. Enter into the 5500A the measured values of each step listed in Table 3-8
when prompted to do so.
5500A
CALIBRATOR
NORMAL
AUX
SCOPE
V,
,
A,
-SENSE,
200V PK
RTD
AUX V
STBY
MAX
OPR
EARTH
SCOPE
HI
7
8
9
1000V
20V
RMS
RMS
MAX
MAX
TRIG
LO
OUT
4
5
6
1V PK
MAX
20V PK
MAX
1
2
3
TC
/ +
0
20V PK
MAX
5500A
Connect the Input leads to the NORMAL output terminals.
Connect the SENSE leads to the AUX terminals.
Figure 3-3. Connections for Calibrating Four-Wire Ohms
Table 3-7. AUX AC Volts Calibration Steps
1.1 V
1.1 V
1.1 V
1.1 V
1.1 V
1.1 V
1.1 V
PREV
BOOST
MENU
µ
dBm
sec
EDIT
SETUP
RESET
m
V
Hz
FIELD
n
W
¡F
NEW
k
A
¡C
CE
REF
p
MEAS
TRIG
M
F
TC
OUT
POWER
MULT
DIV
SHIFT
ENTER
x
÷
I
O
AUX output
300 mV
300 mV
300 mV
3 V
3 V
3 V
3 V
HP3458 4W Ohms
Function
Calibration and Verification
Calibration
Frequency
100 Hz
5 kHz
10 kHz
100 Hz
5 kHz
10 kHz
9.99 Hz
om010f.eps
3
3-9

Advertisement

Table of Contents
loading

Table of Contents