5500A
Service Manual
1-27. AC Current (Non-Sinewave) Specifications
Trianglewave &
Truncated
Sinewave
Ranges [1]
2.9 to 92.999 mA
93 to 929.999 mA
0.93 to 2.19 A
2.2 to 11 A
2.9 to 65.999 mA
66 to 659.999 mA
0.66 to 2.19 A
2.2 to 11 A
[1] All waveforms are p-p output ranges.
[2] Uncertainty is stated in p-p. Amplitude is verified using an rms-responding DMM.
1-30
Uncertainty, tcal ± 5 °C,
Frequency
± (% of output + % of range)
%Output
0.01 to 10 Hz
5.0%
10 to 45 Hz
0.25
45 Hz to 1 kHz
0.25
1 to 10 kHz
0.25
0.01 to 10 Hz
5.0%
10 to 45 Hz
0.25
45 Hz to 1 kHz
0.25
1 to 10 kHz
5.0
10 to 45 Hz
5.0%
45 Hz to 1 kHz
0.5
1 to 5 kHz
5.0
45 to 500 Hz
2.0%
500 Hz to 1 kHz
5.0
0.01 to 10 Hz
5.0 %
10 to 45 Hz
0.25
45 Hz to 1 kHz
0.25
1 to 10 kHz
0.25
0.01 to 10 Hz
5.0%
10 to 45 Hz
0.25
45 Hz to 1 kHz
0.25
1 to 10 kHz
5.0
10 to 45 Hz
5.0%
45 Hz to 1 kHz
0.5
1 to 5 kHz
5.0
45 to 500 Hz
2.0%
500 Hz to 1 kHz
5.0
1-Year Absolute
[2]
%Range
0.5%
0.5
0.25
0.5
0.5%
0.5
0.5
1.0
1.0%
0.5
1.0
0.5%
1.0
0.5%
0.5
0.25
0.5
0.5%
0.5
0.5
1.0
1.0%
0.5
1.0
0.5%
1.0
Maximum
Current
Resolution
Two digits, e.g., 75 mA
Six digits on each range
Two digits
Six digits on each range
Two digits
Six digits on each range
Two digits on each range
Six digits on each range
Two digits, e.g., 50 mA
Six digits on each range
Two digits
Six digits on each range
Two digits
Six digits on each range
Two digits on each range
Six digits on each range