Sc300 Hardware Adjustments - Fluke 5500A Service Manual

Multi-product calibrator
Hide thumbs Also See for 5500A:
Table of Contents

Advertisement

Calibrator
Calibrator
Mainframe
Mainframe
Wave Type
output
(@ 10 kHz)
square
5.0 mV
square
10.9 mV
square
45 mV
square
109 mV
square
0.45V
square
1.09V
square
2.20V
sine
5.0 mV
sine
10.9 mV
sine
45 mV
sine
109 mV
sine
0.45 V
sine
1.09 V
sine
2.20 V
triangle
5.0 mV
triangle
10.9 mV
triangle
45 mV
triangle
109 mV
triangle
0.45 V
triangle
1.09 V
triangle
2.20 V

6-136. SC300 Hardware Adjustments

Before beginning SC300 hardware adjustments, it must determined which
revision of the option is installed in the instrument. To do this, remove the
top cover of the calibrator and look at the circuit board tab protruding
through the guard cover that is closest to the right front corner of the
calibrator. If this tab is marked A4, proceed to the"SC300 Hardware
Adjustments for the A4 Board" section of this manual.
Hardware adjustments must be made to the leveled sine and edge functions each time the
SC300 is repaired. In addition to the adjustment procedures, this section provides lists of
the required equipment and some recommendations on models that have the capabilities
required by these procedures. Equivalent models can be substituted if necessary.
Table 6-71. Wave Generator Verification at 50 Ω
5790A
Conversion
Reading
Factor
(V rms)
2.0000
2.0000
2.0000
2.0000
2.0000
2.0000
2.0000
2.8284
2.8284
2.8284
2.8284
2.8284
2.8284
2.8284
3.4641
3.4641
3.4641
3.4641
3.4641
3.4641
3.4641
Note

SC300 Hardware Adjustments

5790A Reading x
Conversion Factor
(V p-p)
250.00 µV
430.00 µV
1.450 mV
3.370 mV
13.570 mV
32.500 mV
66.100 mV
250.00 µV
430.00 µV
1.450 mV
3.370 mV
13.570 mV
32.500 mV
66.100 mV
250.00 µV
430.00 µV
1.450 mV
3.370 mV
13.570 mV
32.500 mV
66.100 mV
SC300 Option
6
Tolerance
(V p-p)
6-111