Fluke 5500A Service Manual page 202

Multi-product calibrator
Hide thumbs Also See for 5500A:
Table of Contents

Advertisement

5500A
Service Manual
Instrument
Frequency
Counter
Adapter
BNC Cable
AC Measurement
Standard
Adapter
BNC Cable
Spectrum Analyzer
Adapter
BNC Cable
Frequency Counter
BNC Cable
Frequency Counter
BNC Cable
Power Meter
Power Sensor
Power Sensor
30 dB
Reference
Attenuator
Adapter
BNC Cable
6-76
Table 6-41. SC300 Calibration and Verification Equipment (cont.)
Leveled Sine Wave Frequency Verification
PM 6680 with option (PM 9621, PM 9624, or
PM 9625) and (PM 9678)
Pomona #3288
(supplied with SC300)
Leveled Sine Wave Flatness (Low Frequency) Calibration and Verification
Fluke 5790A
with -03 option
Pomona #3288
(supplied with SC300)
Leveled Sine Wave Harmonics Verification
Edge Frequency, AC Voltage Frequency Verification
PM 6680 with option (PM
9678)
(supplied with SC300)
PM 6680
(supplied with SC300)
Leveled Sine Wave Flatness (High Frequency) Calibration and Verification
Hewlett-Packard
E4418A
Hewlett-Packard 8482A
Hewlett-Packard 8481D
Hewlett-Packard
11708A
(supplied with HP
8481D)
Hewlett-Packard
PN 1250-1474
(supplied with SC300)
Model
Range
5 mV p-p to 5.5 V p-p
Frequency
50 kHz to 10 MHz
BNC(f) to Type N(m)
HP 8590A
Pomona #3288
(supplied with SC300)
20 ms to 150 ns, 10 Hz to 10 MHz: < 1.6 ppm
uncertainty
Edge Duty Cycle
Range
-42 to +5.6 dBm
Frequency
10 - 300 MHz
Range
-20 to +19 dBm
Frequency
10 - 300 MHz
Range
-42 to -20 dBm
Frequency
10 - 300 MHz
Range
30 dB
Frequency
50 MHz
BNC(f) to Type N(f)
Minimum Use Specifications
50 kHz to 350 MHz, < 1.6 ppm
uncertainty
BNC(f) to Type N(m)
BNC(f) to Type N(m)