Figure 7-4: Span Cal Menu; Figure 7-5: Flow Cal Menu - Teledyne 465L Operation Manual

Ozone monitor
Table of Contents

Advertisement

Enter Span target
concentration (actual ozone
concentration being supplied
SPAN CAL
ENT
7.2.3.6. Flow Calibration (Optional Electronic Flowmeter only)
The flow calibration allows the instrument to calibrate its internal flowmeter
reading. This should be done with a calibrated flowmeter connected to the "Sample
In" port of the 465L.
Figure 7-5 below shows the Flow Cal menu. After the instrument has stabilized on
the source of span gas, navigate to the Flow Cal menu (See Figure 7-2) and Press
ENT. Next enter the Flow Cal target (the actual flow as measured at the sample
inlet) and press ENT. The display will next show a confirmation menu, ENT TO
CAL. Press ENT to perform the Flow Calibration, or CFG to abort back to the
start. If the calibration is successful, the display will return to the concentration. If
the calibration cannot be performed, an OUT OF RANGE error will be displayed.
Press ENT to confirm and the display will return to the start of the Flow Cal menu.
If the OUT OF RANGE error occurs, it means the Flow Cal cannot be performed
because it would result in an out of range flow slope value for the monitor. This
means that either the flow sensor in the monitor is malfunctioning, causing improper
readings, or the actual flow is different than the target value being entered.
Enter Flow Cal target flow
(actual flow measured at
FLOW CAL
ENT
05509G DCN6806
to monitor)
ENT TO
900
ENT
SPAN
CFG

Figure 7-4: Span Cal Menu

sample inlet)
0800
ENT TO CAL
ENT
CFG

Figure 7-5: Flow Cal Menu

Teledyne API Model 465L O
Teledyne API Model 465L O 3 Monitor Operation Manual
SUCCESSFUL?
ENT
No
OUT OF
RANGE
ENT
SUCCESSFUL?
ENT
No
OUT OF
RANGE
ENT
Monitor Manual
3
RETURN TO
Yes
CONC MODE
RETURN TO
Yes
CONC MODE