9.12.1. Boot Loader – Hardware Setup
Verify that the instrument is configured for an RS-232 connection. See Section 8.2
for details on changing the COM port configuration.
connected to the PC COM port using a 9-pin serial cable (TAPI P/N WR77).
9.12.2. Boot Loader – Software Setup
Table 9-2 lists the required firmware loader software and firmware options.
SOFTWARE TYPE
PC Software
Instrument Firmware
The firmware loader is available on Teledyne API's website at
api.com/software/.
1. Install the firmware loader software onto a PC.
2. Verify the cable connections. Do not power up the instrument.
3. Run the firmware loader application on the PC.
4. Select the COM port that the instrument is connected to.
5. Select the HEX file for the appropriate firmware.
6. Refer to Section 9.12.2.1 to upload the firmware or to Section 9.12.2.2 to
verify the firmware.
72
Table 9-2: Required Software
FILE TYPE/VERSION
FirmwareLoader.EXE
05498-00_XX.HEX (for ppb, ppm, ugm)
05498-01_XX.HEX (for ppb, ppm, pphm)
Teledyne API Model 465L O