Teledyne API Model 465L O 3 Monitor Operation Manual

7.2.4. VIEW Menu

The VIEW menu allows the operator to view various measurement parameters in
real-time. This can be useful for diagnosing various instrument or system problems.
Note that these values are updated in real-time as they are repeatedly scrolled on the
display.
Parameter
MEAS
REF
PRESSURE
LAMP
TEMP
SAMPLE
TEMP
FLOW
*After initial 20 minute warm-up period.

7.2.5. CONFIG Menu

The CONFIG menu displays specific information related to the instrument's
configuration. No changes are made within this menu.
CONFIG
S/N
COMM ADDRESS
STREAM COUNT
REV
BREV
BOOT
48

Table 7-5: VIEW Menu Parameters

Meaning
UV detector reading, measure cycle
UV detector reading, reference cycle mV
Sample pressure
UV lamp temperature
Sample temperature
Sample Gas Flow Rate

Table 7-6: CONFIG Menu Parameters

Teledyne API Model 465L O
Units
mV
psia
ºC
ºC
cc/min
Meaning
Serial number of 465L. (Should match
the serial number tag on the rear panel
of the instrument).
Displays the MODBUS slave ID for
the instrument.
Displays the number of streams
configured. (Always 1 for instruments
without the Multi-stream Option).
Revision of instrument firmware
Revision of sensor firmware
Revision of bootloader firmware
Monitor Manual
3
Normal Range*
250 – 1230
250 – 1230
9.0 – 14.9
50.0 – 52.0
5.0 –50.0
640 – 960
05509G DCN6806