Getting Started; Wiring Requirements For Nema Enclosures - Teledyne 465L Operation Manual

Ozone monitor
Table of Contents

Advertisement

4.

GETTING STARTED

The Model 465L is shipped with the following standard equipment:
Power cord (Rack mount version only)
Instruction manual.
Upon receiving the Model 465L please do the following:
1. Verify that there is no apparent shipping damage. (If damage has occurred
please advise shipper first, then Teledyne API.)
2. When installing the Model 465L, allow a minimum of 4 inches for clearance
at the back of the instrument (rack mount only) and 1 inch of clearance on
each side for proper ventilation.
3. Connect sample inlet line(s) to the sample port on rear panel or bottom panel
of instrument. See Figure 4-2 or Figure 4-3 for rear panel pneumatic
connections.
Connect the exhaust fitting on the rear panel (see Figure 4-2) to a
4. Ensure that there are no communication devices connected until after start-up
is complete.
Disconnect power to the AC mains before making or removing any
A proper earth ground connection must be made to the receptacle
labeled "Earth Ground" on the 3 pin AC connector. Failure to do so
4.1.

Wiring Requirements for NEMA Enclosures

Use appropriate wiring rated for this type of equipment, ensuring that it meets local
and national safety and building requirements.
Ensure that overcurrent protection is used (a 5 A circuit breaker is recommended),
and that it fulfills the following requirements:
• be located as near to the instrument as possible
• quickly and easily accessible
• clearly labeled as the disconnecting device for this instrument
05509G DCN6806
suitable vent outside the monitor area.
WARNING: Electrical Shock Hazard
electrical connections to the 465H.
WARNING: Electrical Shock Hazard
may result in a shock hazard and malfunction of the monitor
Teledyne API Model 465L O
CAUTION
Monitor Manual
3
19

Advertisement

Table of Contents
loading

Table of Contents