Teledyne API Model 465L O 3 Monitor Operation Manual
8. Re-connect power to the instrument.
9. The software portion of the Com port configuration will now automatically be
completed after the next instrument boot-up.

Figure 8-1: Com Configuration Jumper Settings (SW2)

To avoid start-up problems, ensure that there are no devices
attempting communication with the instrument during start-up. Once
the sign-on message is displayed on the front panel, it is safe to begin
54
SW3
1
2
3
4
5
6
7
8
9
ETHERNET
RS485
RS232
Close Indicated Positions Only
SW3
3
5
6
7
8
9
1
2
4
ETHERNET
RS485
RS232
Close Indicated Positions Only
SW3
1 2 3 4 5 6 7
8 9 10
ETHERNET
RS485
RS232
Close Indicated Positions Only
NOTE
communications.
Teledyne API Model 465L O
10
RS232 Configuration
10
RS485 Configuration
Ethernet Configuration
Monitor Manual