9.

MAINTENANCE AND ADJUSTMENTS

9.1.

Maintenance Schedule

Do not look at the UV lamp while the unit is operating. UV light can
cause eye damage. Always use safety glasses made from UV blocking
material. (Generic plastic glasses are not adequate).
Table 9-1 below outlines the suggested maintenance procedures and intervals for
ensuring the 465L continues to operate accurately and reliably. These intervals are
based on continuous (24 hours a day – 7 days a week) operation. These intervals
may be lengthened for intermittent operation.
Maintenance Item
Replace internal
particulate filter
Adjust UV lamp
Replace lamp
Replace sample pump
Replace Sensor Module
Valve
*1
When external sample line pre-filters are used. If pre-filters are not
used, internal sample filter should be replaced every month.
05509G DCN6806
WARNING - RISK OF ELECTRICAL SHOCK
THE OPERATIONS OUTLINED IN THIS CHAPTER ARE TO BE
PERFORMED BY QUALIFIED MAINTENANCE PERSONNEL ONLY!
GENERAL SAFETY HAZARD

Table 9-1: Maintenance Schedule

Recommended Interval
6 months
As Indicated by 'Check Lamp'
LED or status output
As required; when adjustment can
no longer be performed.
2 years
2 years
Teledyne API Model 465L O
CAUTION
*1
Monitor Manual
3
Section
9.3
9.4
9.5
9.6
9.7
63