7.2.3.3. Memory Reset

This function is currently disabled.
7.2.3.4. Zero Calibration
The zero calibration allows the instrument to calibrate its internal ozone offset
factor. This should only be done with a source of zero air connected to the "Sample
In" port of the 465L. Allow the instrument to stabilize on the zero air source before
attempting to zero. This normally takes 10-15 minutes.
When entering the Zero Calibration menu, the prompt "ENT TO CAL" appears on
the display. Simply confirm the calibration by pressing the ENT button to perform
the calibration (to abort the calibration, press and hold the CFG button to return to
the previous menu level.) After pressing ENT, the instrument will automatically
exit the menu mode and return to Concentration mode. The concentration reading
should quickly go to zero. Please note that while measuring zero air, a certain
amount of noise or "dithering" of the concentration about the zero point will occur
and is normal. This noise is typically 1-3 ppb (0.001 - 0.003 ppm) in magnitude.
05509G DCN6806
Teledyne API Model 465L O 3 Monitor Operation Manual
Rear Panel Label
STRM 4 - 3
STRM 5 - 1
STRM 5 – 2
STRM 5 - 3
STRM 6 - 1
STRM 6 – 2
STRM 6 - 3
N/A
N/A
N/A
N/A
N/A
N/A
Teledyne API Model 465L O
Function
STREAM 4 HI-HI ALARM
STREAM 5 ACTIVE
STREAM 5 HI ALARM
STREAM 5 HI-HI ALARM
STREAM 6 ACTIVE
STREAM 6 HI ALARM
STREAM 6 HI-HI ALARM
STREAM 1 INPUT VALVE
STREAM 2 INPUT VALVE
STREAM 3 INPUT VALVE
STREAM 4 INPUT VALVE
STREAM 5 INPUT VALVE
STREAM 6 INPUT VALVE