Teledyne API Model 465L O 3 Monitor Operation Manual
4.3.2. Exhaust Connection
Connect a ¼" OD vent line to the "Exhaust" fitting on the rear panel. This line must
be vented outdoors or to an appropriate discharge system capable of handling ozone.
Local regulations regarding the discharge of ozone to the atmosphere must be
observed. Note that the monitor may be equipped with an internal ozone scrubber
for removal of ozone from the exhaust stream. For safety reasons, the exhaust port
must still be vented appropriately as described above.
Figure 4-6: Rear Panel Electrical I/O Connections
26
Figure 4-7: NEMA Electrical I/O Connections
Teledyne API Model 465L O
Monitor Manual
3
05509G DCN6806