4.4.4. Hi-Current Relay Outputs

Three form C relay outputs are provided on the rear panel on the nine pin connector
(See Figure 4-6). The relays are labeled "Relay 1," "Relay 2," and "Relay 3."
The relays are dry contact type form C (SPDT) relays with Normally Open (N.O.),
Normally Closed (N.C.) and Common (COM) contacts. The relays are capable of
driving loads of up to 250VAC, 5A. For maximum contact life, the relays should
only be used to drive resistive loads. Contact life may be dramatically shortened if
inductive loads are driven without any provision for minimizing high voltage
"inductive kick" that can occur.
These relays have slightly different functions depending on whether the monitor is a
single stream or multi-stream configuration.

Table 4-2: Relay Functions for Single Stream Config

Relay #
1
2
3

Table 4-3: Relay Functions for Multi-Stream Configs

Relay #
1
2
3
* The state of the global alarm represents all the Hi or Hi-
Hi alarm states OR'd together, i.e. if the Hi alarm for one
or more of the sample streams is activated, then the
Global Hi alarm will be activated.
05509G DCN6806
Teledyne API Model 465L O 3 Monitor Operation Manual
Function
System OK Diagnostic
Hi Alarm
Hi-Hi Alarm
Function
System OK Diagnostic
Global Hi Alarm*
Global Hi-Hi Alarm*
Teledyne API Model 465L O
Normal
Operating State
On
Off
Off
Normal
Operating State
On
Off
Off