RS-485

!
CAUTION: DO NOT CONNECT TO FIREWIRE PORTS ON A PC. DAMAGE MAY
OCCUR TO PUMP AND/OR PC.
Mating Face
1
2
3
4
5
6
USB or
USB or
RS-232 IN
P.C.
RS-232
D-Sub 9
D-Sub 9
F/F Cable
NOTE: If a larger chain of pumps with mixed software versions is desired, pumps with software
versions 1.0.4 or lower can be returned to manufacturer for a hardware update.
NOTE: When applying power to pumps in a daisy chain, apply power to the last group in the chain first
and work backwards.
NOTE: When removing power, be aware that power removed from a stand-alone unit will cause power to
be removed from all RS-485 powered units deriving power from that stand-alone unit.
NOTE: Power may be removed from any group of pumps, except the first group, without affecting
communications "downstream"
NOTE: Each pump in the daisy-chain must have a unique address. The first pump must be set to address 00.
508.429.6809
phone
info@kdscientific.com
e-mail
Pin #
RS-485 In Signal
1
PWR IN
2
GND
3
RS-485 A
4
RS-485 B
5
TIED TO PIN 6
**
6
TIED TO PIN 5
**
RS-485 In/Out - IEEE-1394 Sockets
RS-485
PUMP
RS-485
RS-485
OUT
IN
ADDR Ø
IEEE 1394
Cable
Daisy-Chaining via RS-485
508.893.0160
fax
www.kdscientific.com
web
RS-485 Out Signal
Notes
PWR OUT (+30V)
Protected by
*
*
resettable fuse
GND
TIED TO PIN 4
100 Ω, ½ w to
**
**
GND termination
TIED TO PIN 3
**
RS-485 A
RS-485 B
RS-485
RS-485
PUMP
RS-485
• • •
IN
OUT
ADDR 1
IEEE 1394
Cable
KD Scientific Legato 100 Series Syringe Pump User's Manual
Appendices
66
PUMP
ADDR n
Publication 5617-006-REV-2.0