I/O EVENT TRIGGERS

The KDS Legato 110, 111 and 180 pumps permits
easy connection to external I/O devices via a variety of
interfaces, including a built-in 15-pin D-sub connector.
The KDS Legato 110, 111 and 180 Series software
incorporates simple commands for incorporating I/O
controls into your pump Programs. Integrated commands
for control of general trigger signals are integrated in
the software. The following procedure describes the use
of the Advanced Options I/O commands. Following
the procedure is a description of the selection menu
options for each of the I/O command types. Details for
the electrical connection controls and pin-out description
for the 15-pin User I/O Connector are provided in
Appendix C of this manual.
The KDS Legato 110, 111 and 180 manages trigger
events through specific pins on the I/O Connector port
on the back of the pump. The following details define
pin logic:
• Trigger is communicated via Pin 4.
• Event signals are communicated via Pin 2.