Resume Run on Power Loss

If Resume Run on Power Loss is selected the pump will
power up running when power resumes.
NOTE: This feature is only valid in Infuse
Only or Withdraw Only modes. This feature
will only execute if the pump is running
at the time when power is lost, and was
running with no target set.
In the Set-up screen,
select
Resume Run After Power Loss
Accept
the changes made and go back
to the previous screen before Set-up mode
was selected.
Cancel
and go back to the previous screen
before Set-up mode was selected.

Setting and Viewing Pump Information

The information on the pump's parameters are displayed
in the Pump Information Screen.
In the Set-up screen, select
The current information on the pump will be displayed.
• The Model type
• The Software Version
• The Maintenance Due
• The Current Pump Address
Press
Maintenance Reset
maintenance. This will reset it to 100 hours.
Confirm the Maintenance Reset by pressing the
Maintenance Reset
again.
Accept
the changes made by pressing
and go back to the Set-up screen.
Cancel
the changes made by pressing
and go back to the Set-up screen.
508.429.6809
phone
info@kdscientific.com
e-mail
Configuring the KDS Legato 100 Series
Pump Information
to reset the # of hours for
508.893.0160
fax
www.kdscientific.com
web
Press Here
NOTE: Elapsed Timers and Total Infused
Volume will be reset when power resumes.
Counters and timers information prior to
loss of power are lost
Press Here
Press here to
reset the
maintenance
counter
KD Scientific Legato 100 Series Syringe Pump User's Manual
27