Exporting A Program To An
External Computer
Follow these instructions to export a Program to an
external computer.
1. Connect the pump to the external computer via
a USB cable.
2. On the computer, open the HyperTerminal
application and verify that the computer can
communicate with the pump.
3. At the pump, from the Program Main screen,
choose Program Select then choose the Program
you would like to export.
'FILE OPTIONS'
4. Choose the
5. In HyperTerminal, choose Transfer→Capture Text...
6. Enter the File Name that you want to save
the Program under, then choose Start.
7. Choose Export Program to PC.
8. The pump will display a confirmation message in
the Message Area on the screen when the export
is complete. Note that long Programs may take a
few seconds to export.
9. At the destination computer, in HyperTerminal,
when the Program is done transferring, the
message "mend" will appear onscreen.
10. In HyperTerminal, choose Transfer→Capture
Text...→Stop. Your Program is now saved on the
external computer.
NOTE: The KD Scientific Syringe Pump Communicator program, provided on the Legato CD,
can be used to simplify the import / export process for programmable pumps. Please refer to
the Syringe Pump Communicator manual for instructions.
508.429.6809
phone
info@kdscientific.com
e-mail
Operating the KDS Legato 100 Series
IMPORTING A PROGRAM FROM AN EXTERNAL
COMPUTER
Follow these instructions to import a Program from an
external computer.
1. Connect the pump to the external computer via a
2. On the computer, open the HyperTerminal
button.
3. In HyperTerminal, click on the Properties icons and
4. Click on the ASCII Setup button.
5. In the Line Delay field, enter 500. This will set a
6. Click OK twice to return to the HyperTerminal
7. In HyperTerminal, choose Transfer→Send Text
8. Select the Program that you want to import to the
9. Before disconnecting, verify that the Program was
508.893.0160
fax
www.kdscientific.com
web
USB cable and set the Address value to
'00' using the Set Pump Address command in the
Settings menu.
application and verify that the computer can
communicate with the pump.
choose the Settings tab on the pop-up box that is
displayed.
500 millisecond delay after each line is transferred
during the import process.
Main Menu screen.
File...
pump and click Open.
transferred by choosing Program Select from the
Program Main screen at the destination pump,
then browse the Program menu to confirm that the
new Program is now saved at the pump.
KD Scientific Legato 100 Series Syringe Pump User's Manual
49
Publication 5617-006-REV-2.0