Ch.4 Checking the Image Quality

4.1

Correction of Incorporated Densitometer

After completing the installation and settings of the DRYPRO 771, load the films in the supply tray, and print
the test pattern to correct the density.
When a densitometer is available at the facility, use it. If it is not available, use the
one you bought with.
(00)
(01)
(02)
(03)
(04)
(05)
(06)
(07)
(08)
(09)
(10)
S50 DENSITOMETR
/.
INPUT DENSITY
S51 OUT DENSITY
/.
POINT01[000]x0.01D
.
.
.
.
S50 DENSITOMETR
/.
INPUT DENSITY
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
1
Using the Service Maintenance Mode S60, print
out a calibration sheet.
• Bring the Service Maintenance Mode "S60 TEST
PRINT" onto the screen.
• Select "CALIBRATION" using [menu] key, and
press [enter] key.
Print check screen will be displayed following the
print count set up screen.
• Selecting "YES" will register the print queue.
2
Using the density pattern of Service
Maintenance Mode "S50", print out a density
correction sheet.
3
Measure the density with densitometer the
density patch (01) ~ (10).
• Make notes of the measured density beside the
density patch or on the note pad.
4
Select "INPUT DENSITY" of the Service
Maintenance Mode "S50", and press [enter] key.
• Input field for "POINT01" will be displayed.
5
Using the ten key, input the density readings
3
measured in step
.
• Press [
] key of the [menu/select] key after
inputting the value in "POINT01" will switch to the
input field of "POINT02".
6
Upon completion of inputs for all
"POINT01"~"POINT10", press [enter] key.
• The settings will be registered, and returns to the
maintenance menu screen.
• When [exit] key is pressed, a confirmation screen
will be displayed, and prompts to select YES or NO
with the [select] key for registration of the settings.
Data was changed
Save the data?
Selecting "YES" will register the setting, and
returns to the maintenance menu.
Selecting "NO" will return to the maintenance
menu screen without registering the settings.
< 82 >
YES