MESSAGE
Data
Less than 100 test
Memory Low
records can be
generated.
Data Transfer
Bedside unit/
Error
workstation
communication
error.
Enter
High, low, or
Narrower
normal control
Range(s) or
range error.
Min<Max
Min>Max.
Insert Strip
Test strip not fully
appears after
inserted into test
you insert test
strip holder.
strip.
Internal
Faulty internal
Communica-
communications.
tion Error
Internal Error
System Error.
Power Down
Invalid
Conflicting option
Configuration
set, or
configuration
corrupted.
Invalid
Value
Control
>500 mg/dL was
Value
entered.
Invalid CTL
Expiration date not
Code
correct, or CTL code
format invalid.
Invalid
Invalid month or
Expiration
year. Or, date
Date
entered already
expired.
If you cannot resolve a problem, call LifeScan Healthcare Professional Line:
POSSIBLE
CAUSE
Transfer data to free memory
space.
Retry data transfer.
Check ranges. Make sure min
value is less than max.
Remove test strip. When
"Insert Strip" appears, firmly
insert test strip until it stops.
Retry operation. If problem
persists, do not use bedside
unit. Contact LifeScan.
Turn off bedside unit.
Contact LifeScan.
Contact system
administrator to review
configuration settings.
Transfer data.
Re-enter correct control
value.
Check CTL code on control
solution vial and re-enter it.
Check expiration date and
re-enter it.
1 800 524-7226 (USA)
1 888 353-0800 (Canada)