Lifescan SureStep Pro Operator's Manual page 63

Bedside unit professional blood glucose management system
Table of Contents

Advertisement

results marked with the Procedure Err
comment will appear as Proc Err.
• If you select QC/Lin/Unkn,
you may enter an operator
ID and see the results for
the QC, linearity, and
unknown tests a particular
operator has run, or leave
the operator ID field blank
and see the results for all
QC, linearity, and
unknown tests for the last
30 days, beginning with
the most recently performed test.
The control solution level (L, N, H) is listed
next to the QC result. An F following the
solution type indicates the result failed.
UA–UE is listed next to unknown solution
results. L1–L5 is listed next to linearity
solution results. Results for each day are
preceded by a header, which includes the date
and the number of test results for that day.
NOTE
Test results appearing as HIGH indicate
patient, QC, and unknown solution results
greater than 500 mg/dL or linearity solution
results greater than 766 mg/dL.
51