Lifescan SureStep Pro Operator's Manual page 95

Bedside unit professional blood glucose management system
Table of Contents

Advertisement

ERROR
Error 5
The bedside unit
Meter Optics
detected a problem
with the optical
system.
If you cannot resolve a problem, call LifeScan Healthcare Professional Line:
POSSIBLE
CAUSE
1.Clean bedside unit,
2.If you continue to get any
3.Contact LifeScan for help.
1 800 524-7226 (USA)
1 888 353-0800 (Canada)
SOLUTION
especially gray area inside
test strip holder, but
NOTE: Before cleaning,
compare color of test strip
confirmation dot to Color
Chart on test strip bottle. If
glucose level is very high
or very low, take appropri-
ate steps to monitor
patient and provide
medical care.
• Repeat test with new
test strip.
• Make sure test strip
holder is fully inserted into
bedside unit.
• Wait for "Insert Strip" to
appear on display before
inserting test strip.
error message, obtain a
laboratory result. Take
appropriate steps to
monitor patient and
provide medical care while
awaiting laboratory result.
83