Unexpected Messages And Observations - Lifescan SureStep Pro Operator's Manual

Bedside unit professional blood glucose management system
Table of Contents

Advertisement

Unexpected Messages and Observations

MESSAGE
Apply Blood Bedside unit is
ready for blood
application.
Bedside Unit
Bedside unit
Disabled
deleted from
system or
unassigned.
Bedside Unit
Initial data transfer
Needs
is required. Or,
Configuration
configuration is
corrupted.
If you cannot resolve a problem, call LifeScan Healthcare Professional Line:
POSSIBLE
CAUSE
Obtain a drop of capillary
blood and apply it to center
of pink test square.
• If white pad becomes satu-
• If white patches or streaks
Contact system
administrator for current
status of bedside unit.
Transfer data.
1 800 524-7226 (USA)
1 888 353-0800 (Canada)
SOLUTION
rated, you have applied
too much blood. Repeat
application with new test
strip.
appear on confirmation
dot, you have not applied
enough blood. Repeat
application with new test
strip.
85