Quality Control - Lifescan SureStep Pro Operator's Manual

Bedside unit professional blood glucose management system
Table of Contents

Advertisement

Quality Control

Configuration Option
Glucose units
mg/dL
mmol/L
Required QC tests
High
Low
Normal
Control solution
ranges (high, low,
normal) for specified
test strip
Result Display
Numeric Value
Pass/Fail Only
(Pass/Fail available for
SureStepPro 2.0 only)
QC Enforcement
Warning Only
Bedside Unit
lockout
Operator
Lockout
(Time period for
Operator Lockout is
24 hours for
SureStepPro 1.2)
Meaning
The units selected for
displaying and reporting
control and blood glucose
results.
Three glucose controls are
available. Any combination
may be selected.
The passing range for each
control solution level has
been defined. This range
may be the same as or
different from the range
found on the test strip
bottle label.
Determines how quality
control test results are
displayed.
An option to ensure QC is
successfully completed as
scheduled. Warning Only
displays a message but
allows you to perform a
patient test. Bedside Unit
lockout displays the same
message but does not allow
you to perform a patient
test. Operator Lockout
prevents the operator who
has not successfully per-
formed QC testing within
the designated time period
(1 to 365 days) from
performing a patient test.
95