Wipe the test strip holder cover and base with
2
a cotton swab or soft cloth moistened with a
10% bleach solution. Be sure to thoroughly
wipe the area on the inside cover. Clean both
sides of the base. Follow with a swab
moistened with water to remove residual
bleach.
CAUTION
Bleach residue left on the test strip holder may
lead to an error message or an inaccurate, high
result.
Dry the test strip holder with a soft cloth or
3
lint-free tissue. Close the holder and set it aside.
Using a cotton swab or soft cloth moistened
4
with a 10% bleach solution, wipe the lens area
and contact points. Wipe this area even if it does
not appear to be dirty. Use a swab or cloth
moistened with water to remove residual bleach.
Be careful not to scratch the lens area.
contact points
lens area