Troubleshooting
MESSAGE
Invalid
Operator ID format
Operator ID
not correct.
Invalid
Patient ID format
Patient ID
not correct.
Invalid
Phone number not
Phone
entered or more
Number
than four pause
characters entered.
Invalid
Reagent lot
Reagent Lot
number not correct
Number
or field left blank.
Invalid Strip
Strip lot number
Lot Number
not correct. Less
than 10 digits were
entered, or code
out of range.
Low Battery Battery power low. Change batteries.
Modem Not
Bedside unit
Supported by
modem model not
SureStepPro
supported by
System
software.
New
A workstation
Reagent
option is set to
Entry at
allow reagent entry
Workstation
at workstation
Only
only.
No Room for
There are five new
New
(non-validated)
Reagent Lot
reagent lots in
bedside unit list.
If you cannot resolve a problem, call LifeScan Healthcare Professional Line:
88
POSSIBLE
CAUSE
Re-enter your ID using
barcode scanner or
alphanumeric touch screen.
Re-enter patient ID using
barcode scanner or
alphanumeric touch screen.
Re-enter phone number.
Check lot number and
re-enter it.
Check lot number and
re-enter it.
Check with your LifeScan
representative for modem
models supported.
You cannot enter new
reagent from bedside unit.
Use reagent from list, or ask
administrator to allow
reagent entry at bedside
unit.
Transfer data and validate
new reagent lots.
1 800 524-7226 (USA)
1 888 353-0800 (Canada)
SOLUTION