ERROR
Error 3
The bedside unit
Test Strip
detected a problem
with the position of
the test strip.
If you cannot resolve a problem, call LifeScan Healthcare Professional Line:
POSSIBLE
CAUSE
1.Clean bedside unit,
2.If you continue to get any
3.Contact LifeScan for help.
1 800 524-7226 (USA)
1 888 353-0800 (Canada)
SOLUTION
especially contact points in
lens area, but
NOTE: Before cleaning,
compare color of test strip
confirmation dot to Color
Chart on test strip bottle. If
glucose level is very high
or very low, take appropri-
ate steps to monitor
patient and provide
medical care.
• Repeat test with new
test strip.
• Do not move test strip
after you insert it. Remove
test strip only when test is
complete.
error message, obtain a
laboratory result. Take
appropriate steps to
monitor patient and
provide medical care while
awaiting laboratory result.