CHAPTER 9: ALERTS, ALARM & PROFILES
This chapter teaches you about your Dexcom G4 PLATINUM (Pediatric)
CGM System's many alerts and alarm and how to set them.
9.1 SETTING YOUR ALERTS
9.1.1 DEFAULT ALERT/ALARM SETTINGS
The following alerts and alarm are preset on your receiver.
Default Alert/Alarm Settings
What will I see
on the receiver
screen?
What does
What is
this mean?
the default
setting?
On at
High Glucose
200 mg/dL
Alert
Your most
recent sensor
glucose
reading is at or
above the high
alert setting.
Dexcom G4 PLATINUM (Pediatric) System User's Guide
How will the
Will the
receiver notify
receiver
me?
re-notify
me?
Vibrates 2
No, unless
times and then
you have
vibrates/beeps
turned
2 times every
on the
5 minutes until
high alert
confirmed or
snooze
your glucose
feature.
value drops
below the alert
level.
(continued on next page)
|