14
Table 1-B. System Agreement to YSI within CGM
Glucose Ranges (Adult Study)
CGM
Glucose
Number
Range
of paired
mg/dL
CGM-YSI
(mmol/L)
Overall
40-60
(2.2-3.3)
61-80
(3.4-4.4)
81-180
(4.5-10.0)
181-300
(10.1-16.7)
301-350
(16.7-19.4)
351-400
(19.4-22.2)
NOTE: CGM readings are within 40-400 mg/dL (2.2-22.2 mmol/L),
inclusive.
Tables 1-A and 1-B are categorized within CGM glucose ranges. When
you see a CGM reading on your receiver, this table shows you how
likely that reading matches your blood glucose level (measured by YSI in
the study).
Agreement When CGM Reads "LOW" or "HIGH"
The System reports glucose concentrations between 40 and 400
mg/dL (2.2-22.2 mmol/L). When the System determines the glucose
level is below 40 mg/dL (2.2 mmol/L), it displays "LOW" in the Receiver
158
|
Dexcom G4 PLATINUM (Pediatric) Continuous Glucose Monitoring System
Percent
within
15/15%
YSI
9152
71%
512
67%
781
73%
3853
67%
2784
72%
775
82%
447
74%
Percent
Percent
within
within
20/20%
30/30%
YSI
YSI
82%
92%
78%
88%
85%
94%
78%
91%
84%
93%
91%
97%
84%
91%
Percent
Percent
Greater
within
than
40/40%
40/40%
YSI
YSI
97%
3%
94%
6%
98%
2%
97%
3%
96%
4%
98%