14
Table 6-B. Hyperglycemic Alert and Detection Rate
Evaluation in Reference to YSI 15 Minutes Before and
After (Adult Study)
Hyperglycemic
Alert Level
mg/dL (mmol/L)
120 (6.7)
140 (7.8)
180 (10.0)
200 (11.1)
220 (12.2)
240 (13.3)
300 (16.7)
Table 6-C. Hyperglycemic Alert and Detection Rate
Evaluation in Reference to SMBG 30 Minutes Before and
After (Pediatric Study, Ages 2-5 years)
Hyperglycemic
Alert Level
mg/dL (mmol/L)
120 (6.7)
140 (7.8)
180 (10.0)
200 (11.1)
220 (12.2)
240 (13.3)
300 (16.7)
170
|
Dexcom G4 PLATINUM (Pediatric) Continuous Glucose Monitoring System
True
False
Alert
Alert
Rate
Rate
95%
5%
94%
6%
92%
8%
92%
8%
91%
9%
91%
9%
82%
18%
True
False
Alert
Alert
Rate
Rate
92%
8%
90%
10%
87%
13%
85%
15%
81%
19%
80%
20%
71%
29%
Hyperglycemia
Detection Rate
98%
97%
97%
97%
95%
94%
86%
Hyperglycemia
Detection Rate
98%
98%
96%
96%
95%
95%
90%
Hyperglycemia
Missed Detection
Rate
2%
3%
3%
3%
5%
6%
14%
Hyperglycemia
Missed Detection
Rate
2%
2%
4%
4%
5%
5%
10%