Transmitter Electromagnetic Immunity Specifications
(continued from page before)
Immunity Test
Surge
IEC 61000-4-5
Voltage
Dips, Short
Interruptions
and Voltage
Variations on
Power Supply
Input Lines
IEC 61000-4-11
Power
Frequency
(50/60 Hz)
Magnetic Field
IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
Receiver Product Specifications
Part Number
IEC 60601 Test
Level
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
< 5% U
(> 95% dip
T
in U
) for 0.5 cycle
T
40% U
(60% dip in
T
) for 5 cycles
U
T
70% U
(30% dip in
T
) for 25 cycles
U
T
< 5% U
(> 95% dip
T
in U
) for 5 sec
T
3 A/m
MT22608
Dexcom G4 PLATINUM (Pediatric) System User's Guide
Transmitter
Compliance Level
Not applicable
Battery operated
Not applicable
Battery operated
3 A/m
(continued on next page)
Electromagnetic
Environment
Guidance
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial
or hospital
environment.
|
181