16
Table 5-B. Hypoglycemic Alert and Detection Rate
Evaluation in Reference to YSI 15 Minutes Before and
After (Adult Study)
Hypoglycemic
Alert Level
mg/dL (mmol/L)
55 (3.1)
60 (3.3)
70 (3.9)
80 (4.4)
90 (5.0)
Table 5-C. Hypoglycemic Alert and Detection Rate
Evaluation in Reference to SMBG 30 Minutes Before and
After (Pediatric Study, Ages 2-5)
Hypoglycemic
Alert Level
mg/dL (mmol/L)
55 (3.1)
60 (3.3)
70 (3.9)
80 (4.4)
90 (5.0)
100 (5.6)
Hypoglycemia Alert Rate
The Alert Rate shows how often the alert is right or wrong. The True
Alert Rate is the percent of time the device alarmed when the blood
True
False
Alert
Alert
Rate
Rate
50%
50%
64%
36%
79%
21%
87%
13%
90%
10%
True
False
Alert
Alert
Rate
Rate
3%
97%
11%
89%
29%
71%
35%
65%
51%
49%
64%
36%
Dexcom G4 PLATINUM (Pediatric) System User's Guide
Hypoglycemia
Detection Rate
71%
75%
83%
86%
89%
Hypoglycemia
Detection Rate
57%
62%
77%
85%
89%
91%
Hypoglycemia
Missed Detection
Rate
29%
25%
17%
14%
11%
Hypoglycemia
Missed Detection
Rate
43%
38%
23%
15%
11%
9%
|